Q

Name

Description

Content

MV_QAPCON

Name to CONDITIONAL APPROVED

Conditional Approved

MV_QAPQIEP

Before deleting instrument, check if it is in use in Entry Inspection and Process Inspection environment.

2

MV_QAPROV

Name to APPROVED.

Approved

MV_QATUESC

After family range association, it suggests a mandatory calibration as YES

1

MV_QBNDMED

When a calibration is performed, it suggests the first range/standard with non-collected initial value.

1

MV_QCALINC

Uncertainty formula calculation:

1 = (Algebraic Sum (Inc.Tot = Inc. Instr. + IR + RR));

2 = (Square Sum (Inc.Tot = ((Inc.Instr)2 + IR 2 + RR2))

1

MV_QCLPERC

For ranges with uncertainty report with metrological confirmation and percentage calculation type, consider Tolerance/Process field in uncertainty calculation. Options: 1= Yes, performs the calculation by percentage and 2= No, does not perform calculation.

1

MV_QDECPAD

It defines decimal place quantity to be used during data collecting. If content = 1, it searches for decimal place quantity defined in initial Range field. When empty, content of VVC (Standard Registry) field will be considered. When content = 2, it searches decimal place quantity of VVC field.

2

MV_QDIRGRA

Directory

\

MV_QFREQAF

Automatic checking frequency (uses Schumacher Table - cond. of receiving)

Y

MV_QICEWOD

It prints Calibration Certificate in Word® format, being 1=Yes and 2=No

2

MV_QINCDM

It adds Medium Deviance in Total Uncertain.

Y

MV_QINCDP

It considers the highest standard deviation of all measurements of the same range for uncertainty calculation of each range point.

N

MV_QINCRR

It adds the R&R value in uncertainty calculation.

N

MV_QMOESPT

During calibration, it orders ranges/standards by checking type and standard.

1

MV_QMSOMED

It considers uncertainty value of secondary standards/ instrument, used in total uncertainty (1)

If 2, the system does not consider such uncertainties in total uncertainty.

1

MV_QMTIOPC

It indicates calibration screen reopening, being 2=does not reopen the screen after confirmation.

2

MV_QMTISHU

When Schumacher concept is used, it suggests the option extension/reduction of calibration validity as Yes (1).

2

MV_QMTPCES

To Soma type ranges, it considers the decimal places quantity defined in range initial value.

2

MV_QMTDDOT

It defines the name .DOT file.

certprinc.dot

MV_QNGREV

It links the fields that are not considered for instruments revision creation.

QM2_STATUS QM2_BITMAP QM2_CHAVE QM2_ONLINE QM2_BAUDRA QM2_PARITY QM2_PORTA QM2_BITTDAD QM2_STOPS QM2_DESCR

MV_QPATHWD

It defines the path where the original file .DOT is registered.

\SYSTEM\MODELOS

MV_QPATHWT

It defines where file .DOT copy will be recorded.

GETTEMPPATH()

MV_QPDELCA

It allows the calibration maintenance (Alteration/Deletion) with dates inferior to last date of instrument calibration (1=no, 2=yes).

1

MV_QQTDIAS

It defines amount of days to send the e-mail, warning about instrument measurement expiration.

5

MV_QREPRO

Name to REPROVED.

REPROVED

MV_QRRIAPR

Limit value to GOOD report to R&R test.

10.00

MV_QRRICON

Limit value to ACCEPTABLE report to R&R test.

30.00

MV_QRRIREP

Limit value to UNACCEPTABLE report to R&R test.

100.00

MV_QRRTAPR

Showed Text GOOD for R&R test.

GOOD

MV_QRRTCON

Represented text ACCEPTABLE to R&R test.

ACCEPTABLE

MV_QRRTREP

Represented text UNACCEPTABLE to R&R test.

UNACCEPTABLE

MV_QSUGINB

In description, it suggests type B uncertainties of last calibration, as type character (1).

If 2, does not suggest uncertainties.

If instrument does not have calibration, type B uncertainty suggested will be the one previously registered in instruments Family Register.

2

MV_QSULOG

It considers logged user in the system as responsible for instrument calibration.

2

MV_QSUMITB

It defines if type A Uncertain with type B Uncertainty must be summed, to show total Uncertainty.

2

MV_QSUPADS

It defines if secondary/instruments standards used must be suggested in the last review of the main instrument that was used.

2

MV_QVLFIMF

Start/Turn off alert, if final measurement values are out of specified.

Y

MV_QMEDNOC

It considers the average of measurement nominal values, in Tendency graphic.

When parameter is able, that is, when content is equal to 1, the measurement average to reference line printing will be considered. This way, measurement average line will be centered in tendency curve;
When parameter is not able; that is, when content is equal to 2, the System will put the last measurement as reference line (standard configuration).

2

MV_QVLINIF

Start/Turn off alert, if final measurement values are out of specified.

Y

MV_QMTPMOV

It defines movement type to be performed in QMT:

1 = Mov. among Dept of a same branch;

2 = Mov. among Dept of other branches.

1